4.8  Distinguishing Nyquist Criterion and Sampling Theorem

Eq. (4.25) is similar to Fs > 2fmax, posed by the sampling theorem, especially in case one interprets fmax as the bandwidth BW of the signal.

But the reader should distinguish the criterion for zero ISI from the sampling theorem. For example, assume a channel of bandwidth BW = 200 Hz is used to transmit a signal with fmax = BW and the symbol rate is Rsym = 100 bauds. Because Rsym ≤ 2BW is obeyed, a zero ISI operation is possible even with the receiver “sampling” at the symbol rate Rsym, that is, its ADC using FsRx = 100 Hz. However, this sampling frequency is below the required FsRx > 2W = 400 Hz of the sampling theorem. This is not surprising because in digital communications one is not concerned with reconstructing the waveform unambiguously and can afford aliasing. Recall that the goal is “simply” unambiguous detection.11

Decreasing the sampling frequency FsRx reduces the computational cost at the receiver. Hence, a lower-bound on FsRx is of interest and can be obtained as follows: assuming the symbols are independent, at least one signal sample must be used to represent the respective symbol value, such that FsRxRsym. In other words, the minimum FsRx at the receiver corresponds to an oversampling factor of L = 1, i. e., using only one sample to represent a symbol. However, for improved performance, this minimum value is sometimes exceeded in favor of L > 1. Table 4.2 summarizes the discussion.

Table 4.2: Contrasting the Nyquist criterion for zero ISI and the sampling theorem for real-valued signals. It is assumed that the signals are baseband and fmax = BW.
Context Equation Interpretation
Sampling theorem (Eq. (C.30)) Fs > 2BW At least Fs to reconstruct the original signal
Zero-ISI Nyquist criterion Rsym ≤2BW BW Hz allows a maximum of 2BW bauds
Processing at receiver FsRxRsym At least one sample representing a symbol

Table 4.2 can be expanded when one considers complex-valued signals. The alternative definitions of bandwidth discussed in Section E.7.2.0 are important here, together with Eq. (E.39). The results for both real and complex-valued signals are very similar when one considers that an analytic signal of bandwidth BW has a double-sided bandwidth that is half of the one considered for real-valued signals.

Table 4.3: Contrasting the Nyquist criterion for zero ISI and the sampling theorem for analytic complex-valued signals. The signals are baseband with support from 0 to fmax = BW.
Context Equation Interpretation
Sampling theorem (Eq. (E.39)) Fs > BW At least Fs to reconstruct the original signal
Zero-ISI Nyquist criterion RsymBW BW Hz allows a maximum of BW bauds
Processing at receiver FsRxRsym At least one sample representing a symbol

For example, assuming a complex envelope with negligible energy outside the frequency range [0,Fmax] and calling its bandwidth BW = Fmax, Eq. (4.25) can be written as

RsymBW.
(4.34)

For the results in Table 4.3, two things to keep in mind are that each sample or symbol corresponds to two real values and BW is half of the double-sided bandwidth.